Short Guide for Data acquisition reset
not checked - not checked - not checked - not checked
-
stop data acquisition program: ->'System' ->'Exit'
all related windows should disappear, if the 'Cosy-Log' window
stays open
-
terminate or kill all processes which communicate
with the front end electronics
-
xh
-
xh_scaler
-
event_decoding
-
...
-
remove all shared memory sections
(list:'ipcs', remove:'ipcrm -m <ID>')
except the ones with extremely long KEY
e.g. '1127297370'
-
if this doesn't help, kill the process 'commu' owned
by user 'c11'
-
remove tape from Exabyte drive by pressing the button at the front,
write-protect the tape
-
switch several crates off
-
CAMAC crate with controller 'IKPQ02': Mains OFF
-
FASTBUS crate 1: DC OFF
-
FASTBUS crate 2: DC OFF
-
trigger PC: remove mains plug
-
CAMAC crate with controller 'IKPC08' (DRAMS): Mains OFF
-
VME crate : Mains OFF
-
switch several crates on (correct order is essential)
-
CAMAC crate with controller 'IKPQ02': Mains ON
-
FASTBUS crate 1: DC ON
-
FASTBUS crate 2: DC ON
-
trigger PC: insert mains plug
-
CAMAC crate with controller 'IKPC08' (DRAMS): Mains ON
-
VME crate : Mains ON
wait till the displays at the E6 modules show '4'
-
insert a new tape into the Exabyte drive
-
start 'wendy'