C.A.E.N. Technical Information Manual
MOD. C193
32 CHANNEL PROGRAMMABLE DISCRIMINATOR
Table of contents
- 1. DESCRIPTION
- 1.1 FUNCTIONAL DESCRIPTION
- 2. SPECIFICATIONS
- 2.1 PACKAGING
2.2 EXTERNAL COMPONENTS
2.3 CHARACTERISTICS OF THE SIGNAL
2.4 POWER REQUIREMENTS
- 3. CAMAC FUNCTIONS
- 4. OPERATING MODES
- 4.1 LOCAL MODE
4.2 REMOTE MODE
4.3 OUTPUT PULSES WIDTH ADJUSTMENT
- 5. CALIBRATION PROCEDURE
- 5.1 NECESSARY INSTRUMENTS
5.2 PRELIMINARY OPERATIONS
- 6. TEST PROCEDURE
- FUNCTIONAL BLOCK DIAGRAM OF THE MODULE
1. DESCRIPTION
1.1 FUNCTIONAL DESCRIPTION
The model C193 is a 32 channel PROGRAMMABLE DISCRIMINATOR
housed in a 2-unit wide CAMAC unit.
Each channel is NON UPDATING type, and consists of a DAC
connected with a hybrid discriminator.
The module can be programmed either in local mode
(via front panel controls) or in remote mode
(via CAMAC).
An internal 80C32 microprocessor controls both operating
modes.
The module accepts 32 negative differential inputs via the front
panel flat cable connector "IN".
The corresponding outputs are differential ECL and they are
presented at the front panel flat cable connector "OUT".
The output pulses width is adjustable via the front panel
trimmer "WDT" and it does not depend on the over
threshold time.
A front panel connector (AN-OUT) allows the output pulses width
to be monitored.
For each channel, numbered from 0 to 31, the threshold value can be
set in a range from 10 mV to 400 mV (guaranteed) and over
(up to 510 mV), in 2 mV steps.
The selected threshold values are stored into an internal EEPROM.
It is also possible to set the same threshold value for all
channels and to increase/decrease it.
For this purpose, the module is provided with three dummy
channels numbered 92, 93, 94.
The calibration operations are accomplished via CAMAC by using the
CAMAC functions in non-standard mode
(see par. 5).
The functional block diagram of the module is shown in Figure 1.1.
![](CAEN_C193_fig1.gif)
Figure 1.1 Functional Block Diagram of the Module.
2. SPECIFICATIONS
2.1 PACKAGING
2-unit wide CAMAC module.
2.2 EXTERNAL COMPONENTS
- No.1 3-digit segment LED display "THR MON",
showing the threshold value referred to the selected
channel.
- No.1 momentary lever switch "UP-DOWN"
(horizontal writing).
This switch allows the threshold of the selected
channel to be increased/decreased in 2 mV steps.
- No.1 pushbutton "WR".
By pressing this pushbutton the selected threshold value
is stored into the module's internal EEPROM.
- No.1 2-position lever switch "REM".
By setting this switch to the "REM" position,
the module operates in REMOTE mode (to the opposite
position, the LOCAL operating mode is allowed).
- No.1 momentary lever switch "UP-DOWN" (vertical
writing), for the channel number selection.
- No.1 2-digit segment LED display "CH", showing the
selected channel number (0 to 31) or 92, 93, 94
(see par. 4., OPERATING MODES).
- No.1 trimmer "WDT", for the output pulse width
setting.
- No.1 LEMO 00 type connector "AN-OUT", where a voltage
level (~ -6 V to -0.5 V) function of the
output pulse width is available.
- No.1 64-pin fIat cable connector "IN".
Channel inputs.
- No.1 64-pin flat cable connector "OUT".
Channel outputs.
2.3 CHARACTERISTICS OF THE SIGNALS
- INPUTS:
- negative polarity;
- minimum pulse width: 4 nsec;
- double pulse resolution: 8 nsec;
- maximum level: ±5 V;
- maximum frequency: 100 MHz;
- minimum over threshold time: 3 nsec;
- termination: 50 Ohm
(on request, it is possible to change the termination
value);
- OUTPUTS:
- differential ECL;
- pulse width: adjustable in a range from 8 nsec
to 35 nsec;
- termination: 100 Ohm.
- INPUT/OUTPUT DELAY:
2.4 POWER REQUIREMENTS
-6 V at 4.8 A (*)
+6 V at 1.3 A (*)
(*)
Values referred to the total of analog and digital boards,
without output termination. |
3. CAMAC FUNCTIONS
F(0) N A(0-15) |
Reads a channel threshold value,
channel 0 to channel 15 (R1+R8). |
F(0) N+1 A(0-15) |
Reads a channel threshold value,
channel 16 to channel 31 (R1+R8). |
F(16) N A(0-15) |
Writes a channel threshold value,
channel 0 to channel 15 (W1+W8). |
F(16) N+1 A(0-15) |
Writes a channel threshold value,
channel 16 to channel 31 (W1+W8). |
F(17) N A(1) or F(17) N+1 A(1) |
Writes the common threshold value
(for all channels) (W1 +W8). |
F(17) N A(2) or F(17) N+1 A(2) |
Increases the common threshold value
(for all channels) (W1+W8). |
F(17) N A(3) or F(17) N+1 A(3) |
Decreases the common threshold value
(for all channels) (W1+W8). |
F(24) N or F(24) N+1 |
Disables local mode.
After this function has been performed, the module
does not operate in local mode. |
F(26) N or F(26) N+1 |
Enables local mode. |
X response for each valid
function. Q response for each valid function,
unless otherwise specified. |
4. OPERATING MODES
The model C193 can be programmed either in local
(via front panel controls) or in REMOTE mode (via CAMAC).
For both modes, the following paragraphs describe the operations
listed below:
- how to set the threshold value correlated with an
individual channel;
- how to set a common threshold value for all channels.
If it is
necessary, perform the CALIBRATION PROCEDURE
(see par. 5.)
and adjust the output pulse width as required
(see par. 4.3). |
CAUTION:
turn OFF the CAMAC crate before inserting or removing the
module.
4.1.1 THRESHOLD VALUE SETT1NG
(FOR INDIVIDUAL CHANNELS)
- Insert the module into a CAMAC slot.
- Configure the system as required.
- On the module's front panel, set the two-position
lever switch "REM" to the unlabelled
position: the LOCAL operating mode is enabled.
- Turn ON the CAMAC crate.
After a few seconds the "CH" display shows
"0" (number of the current channel) and the
"THR MON" display shows the corresponding
threshold value (mV) previously stored into the internal
EEPROM.
- Let the module warm-up tor at least half an hour.
- By the momentary lever switch "UP-DOWN" located on the
left side of the module (vertical writing), select the number
of the channel for which the new threshold value has to be
defined (on the "CH" display the selected channel
number is shown).
- By the momentary lever switch "UP-DOWN" located on the
module's upper side (horizontal writing), select the required
threshold value.
The threshold value is shown on the "THR MON"
display and the corresponding writing is blinking.
- Press the "WR" pushbutton: after a few seconds, the
selected threshold value is stored into the module's internal
EEPROM and the corresponding writing on the
"THR MON" display comes steady.
- Repeat the procedural steps 6 to 8 for each of the remaining
channels.
4.1.2 COMMON THRESHOLD VALUE SETTING
- If it is necessary, carry out the steps 1 to 5 listed in the
preceding paragraph 4.1.1.
- By the momentary lever switch "UP-DOWN" located on the
left side of the module (vertical writing), increase the number
shown on the "CH" display until it is equal to
"92" (*)
("92" is displayed after "31").
- By the momentary lever switch "THR" located on the
module's upper side, select the required threshold value.
This value (mV) is shown on the "THR MON"
display and the corresponding writing is blinking.
- Press the "WR" pushbutton: after a few seconds, the
selected threshold value is stored into the module's internal
EEPROM and is assumed for all channels.
The corresponding writing on the "THR MON" display
becomes steady and the number "92" is shown on the
"CH" display.
(*)
Three dummy channels are dedicated to the common setting
operations:
"92" -> threshold value setting
"93" -> threshold value decrease
"94" -> threshold value increase.
The set threshold value can be increased or decreased by the value
shown on the "THR MON" display (previously selected
by the quot;THR" switch) by pressing the "WR" push
button after having selected die relevant dummy
channel. |
CAUTION:
turn OFF the CAMAC crate before inserting or removing the
module.
4.2.1 THRESHOLD VALUE SETTING
(FOR INDIVIDUAL CHANNELS)
- Insert the module into a CAMAC slot.
- Configure the System as required.
- On the module's front panel, set the two position lever switch
"REM" to the "REM" position:
the REMOTE operating mode is enabled.
- Turn ON the CAMAC crate: after a few seconds, the "CH"
display shows "0" (number of the current channel)
and the "THR MON" display shows the corresponding
threshold value (mV) previously stored into the internal EEPROM
of the module.
- Let the module warm-up for at least half an hour.
- According to the selected channel number (0-15, 16-31) perform an
F(16) N A(0-15) or an
F(16) N+1 A(n)
(where n=channel number - 16)
CAMAC function with the WRITE Lines W1+W8 set to the required
threshold value (resolution = 2 mV per bit): after a few
seconds the new threshold value is stored into the internal
EEPROM and the two displays "CH" and
"THR MON" show the number of the selected channel
and the new threshold value (mV) respectively.
Any attempt to perform another CAMAC operation before
the new threshold value is shown, will generate X = TRUE and
Q = FALSE |
4.2.2 COMMON THRESHOLD VALUE SETTING
- If it is necessary, carry out the steps 1 to 5 listed in the
preceding paragraph 4.2.1.
- Perform an F(17) N A(1) or an F(17) N+1 A(1)
CAMAC function with the WRITE Lines W1+W8 set to the required
threshold value (resolution = 2 mV per bit):
after a few seconds, the new threshold value is stored into
the module's internal EEPROM, and the two displays "CH"
and "THR MON" show, respectively, "0"
and the set threshold value (mV).
Any attempt to perform another CAMAC operation before
the new threshold value is shown, will generate X = TRUE and
Q = FALSE |
4.2.2.1 COMMON THRESHOLD VALUE DECREASE
- If it is necessary, carry out the steps 1 to 5 listed in the
preceding paragraph 4.2.1.
- Perform an F(17) N A(2) or an F(17) N+1 A(2)
CAMAC function with the WRITE Lines W1+W8 set to the required
decrease value (resolution =2 mV per bit):
after a few seconds the updated threshold values are stored into
the module's internal EEPROM and the two front panel displays
show the "0" channel number and the corresponding
threshold value.
If the set decrease value causes a threshold to be
lower than the minimum allowed value, this threshold will be set to
10 mV (minimum allowed value). |
Any attempt to perform another CAMAC operation before
the new threshold value is shown, will generate X = TRUE and
Q = FALSE |
4.2.2.2 COMMON THRESHOLD VALUE INCREASE
- If it is necessary, carry out the steps 1 to 5 listed in the
preceding paragraph 4.2.1.
- Perform an F(17) N A(3) or an F(17) N+1 A(3)
CAMAC function with the WRITE Lines W1+W8 set to the required
increase value (resolution =2 mV per bit):
after a few seconds the updated threshold values are stored into
the module's internal EEPROM and the two front panel displays show
the "0" channel number and the corresponding threshold
value.
If the set increase value causes a threshold to
exceed the maximum allowed value, this threshold will be set to the
maximum allowed value. |
Any attempt to perform another CAMAC operation before
the new threshold value is shown, will generate X = TRUE and
Q = FALSE |
4.2.3 THRESHOLD VALUE READING
To readout any channel threshold value, carry out the following
operation:
According to the number of the selected channel perform an
F(0) N A(n) or an F(0) N+1 A(n) where
- "N" selects the channels group 0 to 15 with
n=0+15;
- "N+l" selects the channels group 16 to 31 with
n=channel number - 16;
(e.g.: an F(0) N A(15) reads the threshold value from
CH 16; and F(0) N+1 A(15) reads the threshold
value from CH 31).
The threshold value is shown on the R1+R8 READ Lines.
4.2.4 LOCAL MODE ENABLING
To enable the LOCAL operating mode via CAMAC without switching off
the CAMAC crate,
carry out the following operations:
- On the module's front panel, set the "REM" lever
switch to the unlabelled position.
- If the local operating mode had been disabled via CAMAC,
perform an F(26) N or an F(26) N+1 CAMAC
function several times until X and Q responses are TRUE.
4.2.5 LOCAL MODE DISABLING
To disable the LOCAL operating mode, carry out the following
operations:
- On the module's front panel, set the "REM" lever
switch to the "REM" position.
- Perform an F(24) N or an F(24) N+1 CAMAC function
several times until X and Q responses are TRUE.
4.3 OUTPUT PULSES WIDTH ADJUSTMENT
To change the width of the output pulses, carry out the adjustment
operations listed below.
4.3.1 NECESSARY INSTRUMENTS
- No.1 Pulse Generator.
- No.1 Oscilloscope (minimum band width: 100 MHz).
- No.1 NIM-ECL-NIM TRANSLATOR, model CAEN N92.
4.3.2 ADJUSTMENT OPERATIONS
- Connect one of the C193 output channels with one of the eight ECL
input channels of the N92 module.
- Connect the N92 output channel corresponding to the selected input,
with the oscilloscope input
(terminated in 50 Ohm impedance).
- Via Pulse Generator, send an overthreshold input signal to the C193
input channel corresponding to the selected output.
- By a screwdriver, turn the "WDT" trimmer
(located on the C193 front panel) until the required output
pulse width is reached (8 nsec to 35 nsec).
5. CALIBRATION PROCEDURE
Each of the Cl93 channels consists of an 8-bit DAC connected with a
hybrid discriminator circuit.
Since the hybrid circuits differ from one another in the offset and
amplification parameters, it is necessary to calibrate the module
before using it for the first time.
For this reason the module is carefully calibrated before delivering
it to customers.
The calibration operations are enabled (through the module's internal
microprocessor) by inserting the internal jumper "TEST":
in this way, the CAMAC functions act as follows:
F(16) N A(0-15) |
Writes the W1+W8 value into the DACs corresponding to
channels 0 to 15. |
F(l6) N+1 A(0-15) |
Writes the W1+W8 value into the DACs corresponding to
channels 16 to 31. |
F(17) N A(1) |
Formats the EEPROM. By performing this function,
the amplification parameters are set to 1 and the offset
ones are set to 0. |
F(24) N followed by F(16) N A(O-15) |
Stores the amplification parameter corresponding to
channels 0 to 15 (W1+W8). |
F(24) N followed by F(16) N+1 A(O-15) |
Stores the amplification parameter corresponding to
channels 16 to 31 (W1+W8). |
F(26) N followed by F(16) N A(O-15) |
Stores the offset parameter corresponding to channels
0 to 15 (W1+W8). |
F(26) N followed by F(16) N+1 A(O-15) |
Stores the amplification parameter corresponding to
channels 16 to 31 (W1+W8). |
Whenever a threshold value setting operation is accomplished, the
internal software program of the module writes the following value
into the DAC corresponding to the selected channel:
NDAC =
(NTHR - CO) / CA
where: NTHR = CAMAC THRESHOLD,
CO = OFFSET PARAMETER,
CA = AMPLIFICATION PARAMETER.
At the same time, on the "THR" front panel display the
following value is shown:
NDISPLAY = NTHR *2
(-mV)
The calibration operations consist of setting the
CO and
CA values for each channel.
5.1 NECESSARY INSTRUMENTS
- No.1 Oscilloscope (minimum band width: 100 MHz).
- No.1 Pulse Generator capable of producing output pulses having
the following characteristics:
- pulse amplitude: -20 mV to -350 mV;
- pulse duration: 10 nsec;
- frequency: 1 MHz.
- No.1 CAMAC crate provided with a CAMAC controller.
- No.1 NIM-ECL-NIM translator, model CAEN N92.
- No.1 Attenuator module, model CAEN N109 or model CAEN N147A
with a C117B CAMAC interface.
5.2 PRELIMINARY OPERATIONS
- Connect Pulse Generator, Attenuator and Oscilloscope as shown
in the figure below:
![](CAEN_C193_fig2.gif)
- Generate a pulse having the following characteristics:
- width: ~ 10 nsec;
- amplitude: ~ -350 mV;
- rise time/fall time: 2 nsec;
- frequency: 1 MHz.
Assume the pulse amplitude as VH.
- Generate a pulse having the following characteristics:
- width, rise time/fall time, frequency:
same as at step 2;
- amplitude: -30 mV.
Assume the pulse amplitude es VL.
- On the module's printed circuit board, insert the jumper
labelled "TEST": the module is now enabled to
operate in CALIBRATION MODE.
- Connect instruments and modules as shown in the figure below:
![](CAEN_C193_fig3.gif)
The Attenuator's output has to be
connected with the first C193 input channel to be
calibrated. |
- Turn ON instruments and CAMAC crate.
- Via Pulse Generator, generate the VH pulse.
- Via CAMAC Controller, perform several F(l6) N A(0-15) or
F(16) N+l A(CHN°-16)
(according to the selected C193 channel)
with the W1+W8 WRITE Lines set to decreasing values beginning
from 255.
When the W1+W8 value is equal to the discrimination threshold
corresponding to the VH input pulse
(an output signal is shown on the oscilloscope),
assume this value as NdacH.
- Via Pulse Generator, generate the VL pulse.
- Via CAMAC Controller, perform several F(16) N A(0-l5) or
F(l6) N+l A(CHN°-16)
(according to the selected C193 channel)
with the W1+W8 WRITE Lines set to increasing values beginning
from 0.
When the W1+W8 value is equal to the discrimination threshold
corresponding to the VL input pulse
(an output signal is shown on the oscilloscope),
assume this value as NdacL.
- Perform an F(24) N and, after a few seconds, an
F(16) N A(0-l5) or F(16) N+1 A(CHN°-16)
(according to the selected C193 channel)
with the W1+W8 WRITE Lines set to
round (50*CA) where:
CA = (VHVL) / (NdacH-NdacL
round=the nearest integer value of...
After a few seconds, the channel number and the W1+W8 values are
shown on the relevant front panel displays.
THE AMPLIFICATION PARAMETER CORRELATED WITH THE SELECTED CHANNEL
IS NOW STORED INTO THE INTERNAL EEPROM OF THE MODULE.
- Perform an F(26) N and, after a few seconds, an
F(16) N A(0-15) or F(16) N+1 A(CHN°-16)
(according to the selected C193 channel)
with the W1+W8 WRITE Lines set to:
round [(VH/2) - (NdacH *
CA/100)].
If the formula above gives a negative result,
it is necessary to use the following one:
(-1) * { round [(VH/2) - (NdacH *
CA/100)]} + 128 |
After a few seconds, the channel number and the W1+W8 value are
shown on the relevant panel displays.
THE OFFSET PARAMETER CORRELATED WITH THE SELECTED CHANNEL
IS NOW STORED INTO THE INTERNAL EEPROM OF THE MODULE.
- Repeat the procedural steps 7 to 12 for each of the remaining
channels.
- Turn 0FF instruments and crate.
- Remove the "TEST" jumper from the module's printed
circuit board.
THE MODULE IS NOW READY TO OPERATE.
The CA und CO values must
be such as to give
0 <= {NDAC=(NTHR-CO)/CA}
<= 255
If it does not occur, a malfunctioning of the channel is
probable. |
6. TEST PROCEDURE
6.1 INTRODUCTION
The operations to be performed to test the module are listed in the
procedure below and have to be carried out according to their
numerical sequence.
None of the procedural steps can be omitted.
Each procedural step contains the operation to be performed and the
corresponding effect of verification to be accomplished.
The module's station number N corresponds to the slot
in which the edge connector of the module itself is
inserted. |
6.2 NECESSARY INSTRUMENTS
- No.1 Oscilloscope (minimum band width: 100 MHz).
- No.1 Pulse Generator capable of producing pulses with the
following characteristics:
- pulse amplitude: -10 mV to -500 mV;
- pulse duration: 10 nsec;
- maximum frequency: > 100 MHz.
- No.1 CAMAC crate provided with a CAMAC controller.
- No.1 NIM-ECL-NIM translator, model CAEN N92.
- No.1 Attenuator module, model CAEN N109 or model CAEN N147A with
a C117B CAMAC interface.
6.3 PROCEDURE
CAUTION: turn 0FF the CAMAC crate before inserting or
removing the module.
- Connect instruments and modules as shown in the flgure below:
![](CAEN_C193_fig3.gif)
The Attenuator's output has to be
connected with the first C193 input channel to be tested
(CH 0). |
- On the module's front panel, set the two-position lever
switch "REM" to the unlabelled position:
the LOCAL operating mode is enabled.
- Turn ON instruments and CAMAC crate: after a few seconds, on the
module's front panel, the "THR MON" display lights
UP and shows a threshold vaiue (the last value stored into the
internal EEPROM), then the "CH" display shows the
current channel number (0).
- By the momentary lever switch "THR" select a new
threshold value (in 2 mV steps): the "THR MON"
display is blinking.
- Press the "WR" pushbutton: the new threshold value is
stored into the internal EEPROM and the "THR MON"
display writing becomes steady.
- By the momentary lever switch "UP-DOWN"
(vertical writing)
select another channel.
- Repeat the procedural steps 4 to 6 for each of the remaining
channels (up to CH31).
- By the momentary lever switch "UP-DOWN"
(vertical writing)
select the dummy channel 92.
- By the momentary lever switch "THR" select a new common
threshold value: the "THR MON" display is
blinking.
- Push the "WR" pushbutton: after a few seconds the
"THR MON" display writing becomes steady.
- By the momentary lever switch "UP-DOWN"
(vertical writing)
select a channel number 0 to 31 and verify that the selected new
threshold value is shown on the "THR MON" display.
Repeat this operation for each of the remaining channels.
- By the momentary lever switch "UP-DOWN"
(vertical writing)
select the dummy channel 93.
- By the momentary lever switch "THR" select a threshold
decrease value: the "THR MON" display is
blinking.
- Press the "WR" pushbutton: after a few seconds the
new common threshold value is stored into the internal EEPROM
and the "THR MON" display becomes steady.
- By the momentary lever switch "UP-DOWN"
(vertical writing)
verify that the new threshold value is the same for all channels
(0 to 31).
- By the momentary lever switch "UP-DOWN"
(vertical writing)
select the dummy channel 94.
- By the momentary lever switch "THR" select a threshold
increase value: the "THR MON" display is
blinking.
- Press the "WR" pushbutton: after a few seconds, the
new common threshold value is stored into the internal EEPROM
and the "THR MON" display becomes steady.
- By the momentary lever switch "UP-DOWN"
(vertical writing)
verify that the new threshold value is the same for all
channels.
- By the momentary lever switch "UP-DOWN"
(vertical writing)
select one of the 32 discriminator channels (0 to 31).
- Via Pulse Generator, send an overthreshold input signal to
the selected C193 channel.
- With a screwdriver, turn the "WDT" trimmer
(located on the C193 front panel)
and verify that the output pulse displayed on the
Oscilloscope can vary in a range from 8 nsec to
35 nsec.
- Via Pulse Generator, inhibit the input signal.
- On the module's front panel, set the two-position lever switch
"REM" to the "REM" position: the REMOTE
operating mode is enabled.
- According to the selected channel number, perform an
F(16) N A(0-15) or an F(16) N+1 A(n)
(where n=channel number - 16) CAMAC function,
with the W1+W8 WRITE Lines set to a new threshold value
(LSB=2 mV): after a few seconds the selected threshold value
is stored into the internal EEPROM and it is displayed on the
"THR MON" display.
Any attempt to perform another CAMAC
operation before the updated threshold value is shown, will
generate X = TRUE and
Q = FALSE |
- According to the selected channel number, perform an
F(0) N A(0-15) or an F(0) N+1 A(n)
(where n=channel number - 16) CAMAC function:
the set threshold value is read by the R1+R8 READ Lines.
- Via Pulse Generator, send an overthreshold input signal through the
selected C193 channel: on the Oscilloscope, an output pulse is
displayed.
- Repeat the procedural steps 25 to 27 for each of the remaining
channels.
- Perform an F(17) N A(1) or an F(17) N+1 A(1)
CAMAC function with the WRITE Lines W1+W8 set to the required
threshold value (LSB = 2 mV): after a few seconds, the new
common threshold value is stored into the module's internal EEPROM,
and the two displays "CH" and "THR MON"
show "0" and the set threshold value (mV) respectively.
Any
attempt to perform another CAMAC operation before the new
threshold value is shown, will generate X = TRUE and
Q = FALSE |
- According to the selected channel number, perform an
F(0) N A(0-15) or an F(0) N+1 A(n)
(where n=channel number - 16) CAMAC function:
the set threshold value is read by the R1+R8 READ Lines.
- Repeat the procedural step 30 for each of the remaining channels:
the read threshold value must be the same for all channels.
- Perform an F(17) N A(3) or an F(17) N+1 A(3)
CAMAC function with the WRITE Lines W1+W8 set to an increment value
of the common threshold (LSB = 2&nmbsp;mV): after a few seconds,
the updated threshold values are stored into the module's internal
EEPROM and the two front panel displays show the "0"
channel number and the corresponding threshold value.
Any attempt to perform another CAMAC
operation before the updated threshold value is shown, will
generate X = TRUE and
Q = FALSE |
- According to the selected channel number, perform an
F(0) N A(0-15) or an F(0) N+1 A(n)
(where n=channel number - 16) CAMAC function:
the set threshold value is read by the R1+R8 READ Lines.
- Repeat the procedural step 33 for each of the remaining channels:
the read threshold value must be the same for all channels.
- Perform an F(17) N A(2) or an F(17) N+1 A(2)
CAMAC function with the WRITE Lines W1+W8 set to a decrease vaiue
of the common threshold (LSB = 2 mV): after a few seconds, the
updated threshold values are stored into the module's internal
EEPROM and the two front panel displays show the "0"
channel number and the corresponding threshold value.
Any attempt to perform another CAMAC
operation before the updated threshold value is shown, will
generate X = TRUE and
Q = FALSE |
- According to the selected channel number, perform an
F(0) N A(0-15) or an F(0) N+1 A(n)
(where n=channel number - 16) CAMAC function:
the set threshold value is read by the R1+R8 READ Lines.
- Repeat the procedural step 36 for each of the remaining channels:
the read threshold value must be the same for all channels.
- Via Pulse Generator, send an overthreshold input signal through
the selected C193 channel: on the Oscilloscope, an output pulse
is displayed.
- Repeat the preceding procedural step for each of the remaining
channels.
THE MODULE IS NOW TESTED AND OPERATES CORRECTLY.